U.S. Code of Federal Regulations
Regulations most recently checked for updates: Sep 13, 2024
§ 1065.310 - Torque calibration.
(a) Scope and frequency. Calibrate all torque-measurement systems including dynamometer torque measurement transducers and systems upon initial installation and after major maintenance. Use good engineering judgment to repeat the calibration. Follow the torque transducer manufacturer's instructions for linearizing your torque sensor's output. We recommend that you calibrate the torque-measurement system with a reference force and a lever arm.
(b) Recommended procedure to quantify lever-arm length. Quantify the lever-arm length, NIST-traceable within ±0.5% uncertainty. The lever arm's length must be measured from the centerline of the dynamometer to the point at which the reference force is measured. The lever arm must be perpendicular to gravity (i.e., horizontal), and it must be perpendicular to the dynamometer's rotational axis. Balance the lever arm's torque or quantify its net hanging torque, NIST-traceable within ±1% uncertainty, and account for it as part of the reference torque.
(c) Recommended procedure to quantify reference force. We recommend dead-weight calibration, but you may use either of the following procedures to quantify the reference force, NIST-traceable within ±0.5% uncertainty.
(1) Dead-weight calibration. This technique applies a known force by hanging known weights at a known distance along a lever arm. Make sure the weights' lever arm is perpendicular to gravity (i.e., horizontal) and perpendicular to the dynamometer's rotational axis. Apply at least six calibration-weight combinations for each applicable torque-measuring range, spacing the weight quantities about equally over the range. Oscillate or rotate the dynamometer during calibration to reduce frictional static hysteresis. Determine each weight's reference force by multiplying its NIST-traceable mass by the local acceleration of Earth's gravity, as described in § 1065.630. Calculate the reference torque as the weights' reference force multiplied by the lever arm reference length.
(2) Strain gage, load transducer, or proving ring calibration. This technique applies force either by hanging weights on a lever arm (these weights and their lever arm length are not used as part of the reference torque determination) or by operating the dynamometer at different torques. Apply at least six force combinations for each applicable torque-measuring range, spacing the force quantities about equally over the range. Oscillate or rotate the dynamometer during calibration to reduce frictional static hysteresis. In this case, the reference torque is determined by multiplying the force output from the reference meter (such as a strain gage, load transducer, or proving ring) by its effective lever-arm length, which you measure from the point where the force measurement is made to the dynamometer's rotational axis. Make sure you measure this length perpendicular to the reference meter's measurement axis and perpendicular to the dynamometer's rotational axis.
§ 1065.315 - Pressure, temperature, and dewpoint calibration.
(a) Calibrate instruments for measuring pressure, temperature, and dewpoint upon initial installation. Follow the instrument manufacturer's instructions and use good engineering judgment to repeat the calibration, as follows:
(1) Pressure. We recommend temperature-compensated, digital-pneumatic, or deadweight pressure calibrators, with data-logging capabilities to minimize transcription errors. We recommend using calibration reference quantities that are NIST-traceable within ±0.5% uncertainty.
(2) Temperature. We recommend digital dry-block or stirred-liquid temperature calibrators, with data logging capabilities to minimize transcription errors. We recommend using calibration reference quantities for absolute temperature that are NIST-traceable within ±0.5% uncertainty. You may perform linearity verification for temperature measurement systems with thermocouples, RTDs, and thermistors by removing the sensor from the system and using a simulator in its place. Use a NIST-traceable simulator that is independently calibrated and, as appropriate, cold-junction compensated. The simulator uncertainty scaled to absolute temperature must be less than 0.5% of T
(3) Dewpoint. We recommend a minimum of three different temperature-equilibrated and temperature-monitored calibration salt solutions in containers that seal completely around the dewpoint sensor. We recommend using calibration reference quantities for absolute dewpoint temperature that are NIST-traceable within ±0.5% uncertainty.
(b) You may remove system components for off-site calibration. We recommend specifying calibration reference quantities that are NIST-traceable within ±0.5% uncertainty.
§ 1065.301 - Overview and general provisions.
(a) This subpart describes required and recommended calibrations and verifications of measurement systems. See subpart C of this part for specifications that apply to individual instruments.
(b) You must generally use complete measurement systems when performing calibrations or verifications in this subpart. For example, this would generally involve evaluating instruments based on values recorded with the complete system you use for recording test data, including analog-to-digital converters. For some calibrations and verifications, we may specify that you disconnect part of the measurement system to introduce a simulated signal.
(c) If we do not specify a calibration or verification for a portion of a measurement system, calibrate that portion of your system and verify its performance at a frequency consistent with any recommendations from the measurement-system manufacturer, consistent with good engineering judgment.
(d) Use NIST-traceable standards to the tolerances we specify for calibrations and verifications. Where we specify the need to use NIST-traceable standards, you may alternatively use international standards recognized by the CIPM Mutual Recognition Arrangement that are not NIST-traceable.
§ 1065.303 - Summary of required calibration and verifications.
The following table summarizes the required and recommended calibrations and verifications described in this subpart and indicates when these have to be performed:
Table 1 of § 1065.303—Summary of Required Calibration and Verifications
Type of calibration or verification | Minimum frequency |
---|---|
§ 1065.305: Accuracy, repeatability and noise | |
§ 1065.307: Linearity verification | |
§ 1065.308: Continuous gas analyzer system response and updating-recording verification—for gas analyzers not continuously compensated for other gas species | Upon initial installation or after system modification that would affect response. |
§ 1065.309: Continuous gas analyzer system-response and updating-recording verification—for gas analyzers continuously compensated for other gas species | Upon initial installation or after system modification that would affect response. |
§ 1065.310: Torque | Upon initial installation and after major maintenance. |
§ 1065.315: Pressure, temperature, dewpoint | Upon initial installation and after major maintenance. |
§ 1065.320: Fuel flow | Upon initial installation and after major maintenance. |
§ 1065.325: Intake flow | Upon initial installation and after major maintenance. |
§ 1065.330: Exhaust flow | Upon initial installation and after major maintenance. |
§ 1065.340: Diluted exhaust flow (CVS) | Upon initial installation and after major maintenance. |
§ 1065.341: CVS and PFD flow verification (propane check) | Upon initial installation, within 35 days before testing, and after major maintenance. |
§ 1065.342 Sample dryer verification | For thermal chillers: Upon installation and after major maintenance. For osmotic membranes; upon installation, within 35 days of testing, and after major maintenance. |
§ 1065.345: Vacuum leak | For laboratory testing: Upon initial installation of the sampling system, within 8 hours before the start of the first test interval of each duty-cycle sequence, and after maintenance such as pre-filter changes. |
For field testing: After each installation of the sampling system on the vehicle, prior to the start of the field test, and after maintenance such as pre-filter changes. | |
§ 1065.350: CO | Upon initial installation and after major maintenance. |
§ 1065.355: CO NDIR CO | Upon initial installation and after major maintenance. |
§ 1065.360: FID calibration THC FID optimization, and THC FID verification | Calibrate all FID analyzers: upon initial installation and after major maintenance. |
Optimize and determine CH | |
Verify CH | |
Verify C | |
§ 1065.362: Raw exhaust FID O | For all FID analyzers: upon initial installation, and after major maintenance. |
For THC FID analyzers: upon initial installation, after major maintenance, and after FID optimization according to § 1065.360. | |
§ 1065.365: Nonmethane cutter penetration | Upon initial installation, within 185 days before testing, and after major maintenance. |
§ 1065.366: Interference verification for FTIR analyzers | Upon initial installation and after major maintenance. |
§ 1065.369: H | Upon initial installation and after major maintenance. |
§ 1065.370: CLD CO | Upon initial installation and after major maintenance. |
§ 1065.372: NDUV HC and H | Upon initial installation and after major maintenance. |
§ 1065.375: N | Upon initial installation and after major maintenance. |
§ 1065.376: Chiller NO | Upon initial installation and after major maintenance. |
§ 1065.378: NO | Upon initial installation, within 35 days before testing, and after major maintenance. |
§ 1065.390: PM balance and weighing | Independent verification: Upon initial installation, within 370 days before testing, and after major maintenance. |
Zero, span, and reference sample verifications: Within 12 hours of weighing, and after major maintenance. | |
§ 1065.395: Inertial PM balance and weighing | Independent verification: Upon initial installation, within 370 days before testing, and after major maintenance. |
Other verifications: Upon initial installation and after major maintenance. |
§ 1065.305 - Verifications for accuracy, repeatability, and noise.
(a) This section describes how to determine the accuracy, repeatability, and noise of an instrument. Table 1 of § 1065.205 specifies recommended values for individual instruments.
(b) We do not require you to verify instrument accuracy, repeatability, or noise.
However, it may be useful to consider these verifications to define a specification for a new instrument, to verify the performance of a new instrument upon delivery, or to troubleshoot an existing instrument.
(c) In this section we use the letter “y” to denote a generic measured quantity, the superscript over-bar to denote an arithmetic mean (such as y
(d) Conduct these verifications as follows:
(1) Prepare an instrument so it operates at its specified temperatures, pressures, and flows. Perform any instrument linearization or calibration procedures prescribed by the instrument manufacturer.
(2) Zero the instrument as you would before an emission test by introducing a zero signal. Depending on the instrument, this may be a zero-concentration gas, a reference signal, a set of reference thermodynamic conditions, or some combination of these. For gas analyzers, use a zero gas that meets the specifications of § 1065.750.
(3) Span the instrument as you would before an emission test by introducing a span signal. Depending on the instrument, this may be a span-concentration gas, a reference signal, a set of reference thermodynamic conditions, or some combination of these. For gas analyzers, use a span gas that meets the specifications of § 1065.750.
(4) Use the instrument to quantify a NIST-traceable reference quantity, y
(5) Sample and record values for 30 seconds (you may select a longer sampling period if the recording update frequency is less than 0.5 Hz), record the arithmetic mean, y
(6) Also, if the reference quantity is not absolutely constant, which might be the case with a reference flow, sample and record values of y
(7) Subtract the reference value, y
(8) Repeat the steps specified in paragraphs (d)(2) through (7) of this section until you have ten arithmetic means (y
(9) Use the following values to quantify your measurements:
(i) Accuracy. Instrument accuracy is the absolute difference between the reference quantity, y
(ii) Repeatability. Repeatability is two times the standard deviation of the ten errors (that is, repeatability = 2 ·
(iii) Noise. Noise is two times the root-mean-square of the ten standard deviations (that is, noise = 2 · rms
(10) You may use a measurement instrument that does not meet the accuracy, repeatability, or noise specifications in Table 1 of § 1065.205, as long as you meet the following criteria:
(i) Your measurement systems meet all the other required calibration, verification, and validation specifications that apply as specified in the regulations.
(ii) The measurement deficiency does not adversely affect your ability to demonstrate compliance with the applicable standards in this chapter.
§ 1065.307 - Linearity verification.
(a) Scope and frequency. Perform linearity verification on each measurement system listed in Table 1 of this section at least as frequently as indicated in Table 1 of § 1065.303, consistent with measurement system manufacturer's recommendations and good engineering judgment. The intent of linearity verification is to determine that a measurement system responds accurately and proportionally over the measurement range of interest. Linearity verification generally consists of introducing a series of at least 10 reference values to a measurement system. The measurement system quantifies each reference value. The measured values are then collectively compared to the reference values by using a least-squares linear regression and the linearity criteria specified in Table 1 of this section.
(b) Performance requirements. If a measurement system does not meet the applicable linearity criteria referenced in Table 1 of this section, correct the deficiency by re-calibrating, servicing, or replacing components as needed. Repeat the linearity verification after correcting the deficiency to ensure that the measurement system meets the linearity criteria. Before you may use a measurement system that does not meet linearity criteria, you must demonstrate to us that the deficiency does not adversely affect your ability to demonstrate compliance with the applicable standards in this chapter.
(c) Procedure. Use the following linearity verification protocol, or use good engineering judgment to develop a different protocol that satisfies the intent of this section, as described in paragraph (a) of this section:
(1) In this paragraph (c), the letter “y” denotes a generic measured quantity, the superscript over-bar denotes an arithmetic mean (such as y
(2) Use good engineering judgment to operate a measurement system at normal operating conditions. This may include any specified adjustment or periodic calibration of the measurement system.
(3) If applicable, zero the instrument as you would before an emission test by introducing a zero signal. Depending on the instrument, this may be a zero-concentration gas, a reference signal, a set of reference thermodynamic conditions, or some combination of these. For gas analyzers, use a zero gas that meets the specifications of § 1065.750 and introduce it directly at the analyzer port.
(4) If applicable, span the instrument as you would before an emission test by introducing a span signal. Depending on the instrument, this may be a span-concentration gas, a reference signal, a set of reference thermodynamic conditions, or some combination of these. For gas analyzers, use a span gas that meets the specifications of § 1065.750 and introduce it directly at the analyzer port.
(5) If applicable, after spanning the instrument, check zero with the same signal you used in paragraph (c)(3) of this section. Based on the zero reading, use good engineering judgment to determine whether or not to rezero and or re-span the instrument before continuing.
(6) For all measured quantities, use the instrument manufacturer's recommendations and good engineering judgment to select reference values, y
(7) Use the instrument manufacturer's recommendations and good engineering judgment to select the order in which you will introduce the series of reference values. For example, you may select the reference values randomly to avoid correlation with previous measurements and to avoid hysteresis; you may select reference values in ascending or descending order to avoid long settling times of reference signals; or you may select values to ascend and then descend to incorporate the effects of any instrument hysteresis into the linearity verification.
(8) Generate reference quantities as described in paragraph (d) of this section. For gas analyzers, use gas concentrations known to be within the specifications of § 1065.750 and introduce them directly at the analyzer port.
(9) Introduce a reference signal to the measurement instrument.
(10) Allow time for the instrument to stabilize while it measures the value at the reference condition. Stabilization time may include time to purge an instrument and time to account for its response.
(11) At a recording frequency of at least f Hz, specified in Table 1 of § 1065.205, measure the value at the reference condition for 30 seconds (you may select a longer sampling period if the recording update frequency is less than 0.5 Hz) and record the arithmetic mean of the recorded values, y
(12) Repeat the steps in paragraphs (c)(9) though (11) of this section until measurements are complete at each of the reference conditions.
(13) Use the arithmetic means, y
(d) Reference signals. This paragraph (d) describes recommended methods for generating reference values for the linearity-verification protocol in paragraph (c) of this section. Use reference values that simulate actual values, or introduce an actual value and measure it with a reference-measurement system. In the latter case, the reference value is the value reported by the reference-measurement system. Reference values and reference-measurement systems must be NIST-traceable. We recommend using calibration reference quantities that are NIST-traceable within ±0.5% uncertainty, if not specified elsewhere in this part 1065. Use the following recommended methods to generate reference values or use good engineering judgment to select a different reference:
(1) Speed. Run the engine or dynamometer at a series of steady-state speeds and use a strobe, photo tachometer, or laser tachometer to record reference speeds.
(2) Torque. Use a series of calibration weights and a calibration lever arm to simulate engine torque. You may instead use the engine or dynamometer itself to generate a nominal torque that is measured by a reference load cell or proving ring in series with the torque-measurement system. In this case, use the reference load cell measurement as the reference value. Refer to § 1065.310 for a torque-calibration procedure similar to the linearity verification in this section.
(3) Electrical power, current, and voltage. You must perform linearity verification for either electrical power meters, or for current and voltage meters. Perform linearity verifications using a reference meter and controlled sources of current and voltage. We recommend using a complete calibration system that is suitable for the electrical power distribution industry.
(4) Fuel and DEF mass flow rate. Use a gravimetric reference measurement (such as a scale, balance, or mass comparator) and a container. Use a stopwatch or timer to measure the time intervals over which reference masses of fluid pass through the mass flow rate meter. Use good engineering judgment to correct the reference mass flowing through the mass flow rate meter for buoyancy effects from any tubes, temperature probes, or objects submerged in the fluid in the container that are not attached to the container. If the container has any tubes or wires connected to the container, recalibrate the gravimetric reference measurement device with them connected and at normal operating pressure using calibration weights that meet the requirements in § 1065.790. The corrected reference mass that flowed through the mass flow rate meter during a time interval divided by the duration of the time interval is the average reference mass flow rate. For meters that report a different quantity (such as actual volume, standard volume, or moles), convert the reported quantity to mass. For meters that report a cumulative quantity calculate the average measured mass flow rate as the difference in the reported cumulative mass during the time interval divided by the duration of the time interval. For measuring flow rate of gaseous fuel prevent condensation on the fuel container and any attached tubes, fittings, or regulators.
(5) Flow rates—inlet air, dilution air, diluted exhaust, raw exhaust, or sample flow. Use a reference flow meter with a blower or pump to simulate flow rates. Use a restrictor, diverter valve, a variable-speed blower or a variable-speed pump to control the range of flow rates. Use the reference meter's response as the reference values.
(i) Reference flow meters. Because the flow range requirements for these various flows are large, we allow a variety of reference meters. For example, for diluted exhaust flow for a full-flow dilution system, we recommend a reference subsonic venturi flow meter with a restrictor valve and a blower to simulate flow rates. For inlet air, dilution air, diluted exhaust for partial-flow dilution, raw exhaust, or sample flow, we allow reference meters such as critical flow orifices, critical flow venturis, laminar flow elements, master mass flow standards, or Roots meters. Make sure the reference meter is calibrated and its calibration is NIST-traceable. If you use the difference of two flow measurements to determine a net flow rate, you may use one of the measurements as a reference for the other.
(ii) Reference flow values. Because the reference flow is not absolutely constant, sample and record values of n
(6) Gas division. Use one of the two reference signals:
(i) At the outlet of the gas-division system, connect a gas analyzer that meets the linearity verification described in this section and has not been linearized with the gas divider being verified. For example, verify the linearity of an analyzer using a series of reference analytical gases directly from compressed gas cylinders that meet the specifications of § 1065.750. We recommend using a FID analyzer or a PMD or MPD O
(ii) Using good engineering judgment and the gas divider manufacturer's recommendations, use one or more reference flow meters to measure the flow rates of the gas divider and verify the gas-division value.
(7) Continuous constituent concentration. For reference values, use a series of gas cylinders of known gas concentration containing only a single constituent of interest with balance of purified air or purified N
(8) Temperature. You may perform the linearity verification for temperature measurement systems with thermocouples, RTDs, and thermistors by removing the sensor from the system and using a simulator in its place. Use a NIST-traceable simulator that is independently calibrated and, as appropriate, cold-junction-compensated. The simulator uncertainty scaled to absolute temperature must be less than 0.5% of T
(9) Mass. For linearity verification for gravimetric PM balances, fuel mass scales, and DEF mass scales, use external calibration weights that meet the requirements in § 1065.790. Perform the linearity verification for fuel mass scales and DEF mass scales with the in-use container, installing all objects that interface with the container. For example, this includes all tubes, temperature probes, and objects submerged in the fluid in the container; it also includes tubes, fittings, regulators, and wires, and any other objects attached to the container. We recommend that you develop and apply appropriate buoyancy corrections for the configuration of your mass scale during normal testing, consistent with good engineering judgment. Account for the scale weighing a calibration weight instead of fluid if you calculate buoyancy corrections. You may also correct for the effect of natural convection currents from temperature differences between the container and ambient air. Prepare for linearity verification by taking the following steps for vented and unvented containers:
(i) If the container is vented to ambient, fill the container and tubes with fluid above the minimum level used to trigger a fill operation; drain the fluid down to the minimum level; tare the scale; and perform the linearity verification.
(ii) If the container is rigid and not vented, drain the fluid down to the minimum level; fill all tubes attached to the container to normal operating pressure; tare the scale; and perform the linearity verification.
(e) Measurement systems that require linearity verification. Table 1 of this section indicates measurement systems that require linearity verification, subject to the following provisions:
(1) Perform linearity verification more frequently based on the instrument manufacturer's recommendation or good engineering judgment.
(2) The expression “x
(3) The expression “max” generally refers to the absolute value of the reference value used during linearity verification that is furthest from zero. This is the value used to scale the first and third tolerances in Table 1 of this section using a
(i) For linearity verification of a PM balance, m
(ii) For linearity verification of a torque measurement system used with the engine's primary output shaft, T
(iii) For linearity verification of a fuel mass scale, m
(iv) For linearity verification of a DEF mass scale, m
(v) For linearity verification of a fuel flow rate meter, m
(vi) For linearity verification of a DEF flow rate meter, m
(vii) For linearity verification of an intake-air flow rate meter, n
(viii) For linearity verification of a raw exhaust flow rate meter, n
(ix) For linearity verification of an electrical-power measurement system used to determine the engine's primary output shaft torque, P
(x) For linearity verification of an electrical-current measurement system used to determine the engine's primary output shaft torque, I
(xi) For linearity verification of an electrical-voltage measurement system used to determine the engine's primary output shaft torque, V
(4) The specified ranges are inclusive. For example, a specified range of 0.98-1.02 for a
(5) Table 2 of this section describes optional verification procedures you may perform instead of linearity verification for certain systems. The following provisions apply for the alternative verification procedures:
(i) Perform the propane check verification described in § 1065.341 at the frequency specified in Table 1 of § 1065.303.
(ii) Perform the carbon balance error verification described in § 1065.543 on all test sequences that use the corresponding system. It must also meet the restrictions listed in Table 2 of this section. You may evaluate the carbon balance error verification multiple ways with different inputs to validate multiple flow-measurement systems.
(6) You must meet the a
(7) Linearity verification is required for the following temperature measurements:
(i) The following temperature measurements always require linearity verification:
(A) Air intake.
(B) Aftertreatment bed(s), for engines tested with aftertreatment devices subject to cold-start testing.
(C) Dilution air for gaseous and PM sampling, including CVS, double-dilution, and partial-flow systems.
(D) PM sample.
(E) Chiller sample, for gaseous sampling systems that use thermal chillers to dry samples and use chiller temperature to calculate the dewpoint at the outlet of the chiller. For your testing, if you choose to use a high alarm temperature setpoint for the chiller temperature as a constant value in determining the amount of water removed from the emission sample, you may use good engineering judgment to verify the accuracy of the high alarm temperature setpoint instead of linearity verification on the chiller temperature. To verify that the alarm trip point value is no less than 2.0 °C below the reference value at the trip point, we recommend that you input a reference simulated temperature signal below the alarm trip point and increase this signal until the high alarm trips.
(F) Transmission oil.
(G) Axle gear oil.
(ii) Linearity verification is required for the following temperature measurements if these temperature measurements are specified by the engine manufacturer:
(A) Fuel inlet.
(B) Air outlet to the test cell's charge air cooler air outlet, for engines tested with a laboratory heat exchanger that simulates an installed charge air cooler.
(C) Coolant inlet to the test cell's charge air cooler, for engines tested with a laboratory heat exchanger that simulates an installed charge air cooler.
(D) Oil in the sump/pan.
(E) Coolant before the thermostat, for liquid-cooled engines.
(8) Linearity verification is required for the following pressure measurements:
(i) The following pressure measurements always require linearity verification:
(A) Air intake restriction.
(B) Exhaust back pressure as required in § 1065.130(h).
(C) Barometer.
(D) CVS inlet gage pressure where the raw exhaust enters the tunnel.
(E) Sample dryer, for gaseous sampling systems that use either osmotic-membrane or thermal chillers to dry samples. For your testing, if you choose to use a low alarm pressure setpoint for the sample dryer pressure as a constant value in determining the amount of water removed from the emission sample, you may use good engineering judgment to verify the accuracy of the low alarm pressure setpoint instead of linearity verification on the sample dryer pressure. To verify that the trip point value is no more than 4.0 kPa above the reference value at the trip point, we recommend that you input a reference pressure signal above the alarm trip point and decrease this signal until the low alarm trips.
(ii) Linearity verification is required for the following pressure measurements if these pressure measurements are specified by the engine manufacturer:
(A) The test cell's charge air cooler and interconnecting pipe pressure drop, for turbo-charged engines tested with a laboratory heat exchanger that simulates an installed charge air cooler.
(B) Fuel outlet.
(f) Performance criteria for measurement systems. Table 1 follows:
Table 1 of § 1065.307—Measurement Systems That Require Linearity Verification
Measurement system | Quantity | Linearity criteria | |||
---|---|---|---|---|---|
| | |||||
Speed | ≤0.05% · | 0.98-1.02 | ≤2% · | ≥0.990 | |
Torque | ≤1% · | 0.98-1.02 | ≤2% · | ≥0.990 | |
Electrical power | ≤1% · | 0.98-1.02 | ≤2% · | ≥0.990 | |
Current | ≤1% · | 0.98-1.02 | ≤2% · | ≥0.990 | |
Voltage | ≤1% · | 0.98-1.02 | ≤2% · | ≥0.990 | |
Fuel flow rate | ≤1% · | 0.98-1.02 | ≤2% · | ≥0.990 | |
Fuel mass scale | ≤0.3% · | 0.996-1.004 | ≤0.4% · | ≥0.999 | |
DEF flow rate | ≤1% · | 0.98-1.02 | ≤2% · | ≥0.990 | |
DEF mass scale | ≤0.3% · | 0.996-1.004 | ≤0.4% · | ≥0.999 | |
Intake-air flow rate a | ≤1% · | 0.98-1.02 | ≤2% · | ≥0.990 | |
Dilution air flow rate a | ≤1% · | 0.98-1.02 | ≤2% · | ≥0.990 | |
Diluted exhaust flow rate a | ≤1% · | 0.98-1.02 | ≤2% · | ≥0.990 | |
Raw exhaust flow rate a | ≤1% · | 0.98-1.02 | ≤2% · | ≥0.990 | |
Batch sampler flow rates a | ≤1% · | 0.98-1.02 | ≤2% · | ≥0.990 | |
Gas dividers | ≤0.5% · | 0.98-1.02 | ≤2% · | ≥0.990 | |
Gas analyzers for laboratory testing | ≤0.5% · | 0.99-1.01 | ≤1% · | ≥0.998 | |
Gas analyzers for field testing | ≤1% · | 0.99-1.01 | ≤1% · | ≥0.998 | |
Electrical aerosol analyzer for field testing | ≤5% · | 0.85-1.15 | ≤10% · | ≥0.950 | |
Photoacoustic analyzer for field testing | ≤5% · | 0.90-1.10 | ≤10% · | ≥0.980 | |
PM balance | ≤1% · | 0.99-1.01 | ≤1% · | ≥0.998 | |
Pressures | ≤1% · p | 0.99-1.01 | ≤1% · | ≥0.998 | |
Dewpoint for intake air, PM-stabilization and balance environments | ≤0.5% · | 0.99-1.01 | ≤0.5% · | ≥0.998 | |
Other dewpoint measurements | ≤1% · | 0.99-1.01 | ≤1% · | ≥0.998 | |
Analog-to-digital conversion of temperature signals | ≤1% · | 0.99-1.01 | ≤1% · | ≥0.998 |
a For flow meters that determine volumetric flow rate,
(g) Alternative verification procedures. Table 2 follows:
Table 2 of § 1065.307—Optional Verification to Linearity Verification
Measurement system | § 1065.341 | § 1065.543 | Restrictions for § 1065.543 |
---|---|---|---|
Intake-air flow rate | Yes | Yes | Determine raw exhaust flow rate using the intake-air flow rate signal as an input into Eq. 1065.655-24 and determine mass of CO |
Dilution air flow rate for CVS | Yes | No | Not allowed. |
Diluted exhaust flow rate for CVS | Yes | Yes | Determine mass of CO |
Raw exhaust flow rate for exhaust stack | Yes | Yes | Determine mass of CO |
Flow measurements in a PFD (usually dilution air and diluted exhaust streams) used to determine the dilution ratio in the PFD | Yes | Yes | Determine mass of CO |
Batch sampler flow rates | Yes | No | Not allowed. |
Fuel mass flow rate | No | Yes | Determine mass of a carbon-carrying fluid stream used as an input into Eq. 1065.643-1 using the fuel mass flow rate meter. |
Fuel mass scale | No | Yes | Determine mass of a carbon-carrying fluid stream used as an input into Eq. 1065.643-1 using the fuel mass scale. |
§ 1065.308 - Continuous gas analyzer system-response and updating-recording verification—for gas analyzers not continuously compensated for other gas species.
(a) Scope and frequency. This section describes a verification procedure for system response and updating-recording frequency for continuous gas analyzers that output a gas species mole fraction (i.e., concentration) using a single gas detector, i.e., gas analyzers not continuously compensated for other gas species measured with multiple gas detectors. See § 1065.309 for verification procedures that apply to continuous gas analyzers that are continuously compensated for other gas species measured with multiple gas detectors. Perform this verification to determine the system response of the continuous gas analyzer and its sampling system. This verification is required for continuous gas analyzers used for transient or ramped-modal testing. You need not perform this verification for batch gas analyzer systems or for continuous gas analyzer systems that are used only for discrete-mode testing. Perform this verification after initial installation (i.e., test cell commissioning) and after any modifications to the system that would change system response. For example, perform this verification if you add a significant volume to the transfer lines by increasing their length or adding a filter; or if you reduce the frequency at which the gas analyzer updates its output or the frequency at which you sample and record gas-analyzer concentrations.
(b) Measurement principles. This test verifies that the updating and recording frequencies match the overall system response to a rapid change in the value of concentrations at the sample probe. Gas analyzers and their sampling systems must be optimized such that their overall response to a rapid change in concentration is updated and recorded at an appropriate frequency to prevent loss of information. This test also verifies that the measurement system meets a minimum response time. You may use the results of this test to determine transformation time, t
(c) System requirements. Demonstrate that each continuous analyzer has adequate update and recording frequencies and has a minimum rise time and a minimum fall time during a rapid change in gas concentration. You must meet one of the following criteria:
(1) The product of the mean rise time, t
(2) The frequency at which the system records an updated concentration must be at least 5 Hz. This criterion assumes that the frequency content of significant changes in emission concentrations during emission testing do not exceed 1 Hz. Also, the mean rise time must be at or below 10 seconds and the mean fall time must be at or below 10 seconds.
(3) You may use other criteria if we approve the criteria in advance.
(4) You may meet the overall PEMS verification in § 1065.920 instead of the verification in this section for field testing with PEMS.
(d) Procedure. Use the following procedure to verify the response of each continuous gas analyzer:
(1) Instrument setup. Follow the analyzer manufacturer's start-up and operating instructions. Adjust the measurement system as needed to optimize performance. Run this verification with the analyzer operating in the same manner you will use for emission testing. If the analyzer shares its sampling system with other analyzers, and if gas flow to the other analyzers will affect the system response time, then start up and operate the other analyzers while running this verification test. You may run this verification test on multiple analyzers sharing the same sampling system at the same time. If you use any analog or real-time digital filters during emission testing, you must operate those filters in the same manner during this verification.
(2) Equipment setup. We recommend using minimal lengths of gas transfer lines between all connections and fast-acting three-way valves (2 inlets, 1 outlet) to control the flow of zero and blended span gases to the sample system's probe inlet or a tee near the outlet of the probe. If you inject the gas at a tee near the outlet of the probe, you may correct the transformation time, t
(3) Data collection. (i) Start the flow of zero gas.
(ii) Allow for stabilization, accounting for transport delays and the slowest analyzer's full response.
(iii) Start recording data. For this verification you must record data at a frequency greater than or equal to that of the updating-recording frequency used during emission testing. You may not use interpolation or filtering to alter the recorded values.
(iv) Switch the flow to allow the blended span gases to flow to the analyzer. If you intend to use the data from this test to determine t
(v) Allow for transport delays and the slowest analyzer's full response.
(vi) Switch the flow to allow zero gas to flow to the analyzer. If you intend to use the data from this test to determine t
(vii) Allow for transport delays and the slowest analyzer's full response.
(viii) Repeat the steps in paragraphs (d)(3)(iv) through (vii) of this section to record seven full cycles, ending with zero gas flowing to the analyzers.
(ix) Stop recording.
(e) Performance evaluation. (1) If you choose to demonstrate compliance with paragraph (c)(1) of this section, use the data from paragraph (d)(3) of this section to calculate the mean rise time, t
(2) If a measurement system fails the criterion in paragraph (e)(1) of this section, ensure that signals from the system are updated and recorded at a frequency of at least 5 Hz. In no case may the mean rise time or mean fall time be greater than 10 seconds.
(3) If a measurement system fails the criteria in paragraphs (e)(1) and (2) of this section, you may use the measurement system only if the deficiency does not adversely affect your ability to show compliance with the applicable standards in this chapter.
(f) Transformation time, t
(g) Optional procedure. Instead of using a three-way valve to switch between zero and span gases, you may use a fast-acting two-way valve to switch sampling between ambient air and span gas at the probe inlet. For this alternate procedure, the following provisions apply:
(1) If your probe is sampling from a continuously flowing gas stream (e.g., a CVS tunnel), you may adjust the span gas flow rate to be different than the sample flow rate.
(2) If your probe is sampling from a gas stream that is not continuously flowing (e.g., a raw exhaust stack), you must adjust the span gas flow rate to be less than the sample flow rate so ambient air is always being drawn into the probe inlet. This avoids errors associated with overflowing span gas out of the probe inlet and drawing it back in when sampling ambient air.
(3) When sampling ambient air or ambient air mixed with span gas, all the analyzer readings must be stable within ±0.5% of the target gas concentration step size. If any analyzer reading is outside the specified range, you must resolve the problem and verify that all the analyzer readings meet this specification.
(4) For oxygen analyzers, you may use purified N
§ 1065.309 - Continuous gas analyzer system-response and updating-recording verification—for gas analyzers continuously compensated for other gas species.
(a) Scope and frequency. This section describes a verification procedure for system response and updating-recording frequency for continuous gas analyzers that output a single gas species mole fraction (i.e., concentration) based on a continuous combination of multiple gas species measured with multiple detectors (i.e., gas analyzers continuously compensated for other gas species). See § 1065.308 for verification procedures that apply to continuous gas analyzers that are not continuously compensated for other gas species or that use only one detector for gaseous species. Perform this verification to determine the system response of the continuous gas analyzer and its sampling system. This verification is required for continuous gas analyzers used for transient or ramped-modal testing. You need not perform this verification for batch gas analyzers or for continuous gas analyzers that are used only for discrete-mode testing. For this check we consider water vapor a gaseous constituent. This verification does not apply to any processing of individual analyzer signals that are time-aligned to their t
(b) Measurement principles. This procedure verifies that the updating and recording frequencies match the overall system response to a rapid change in the value of concentrations at the sample probe. It indirectly verifies the time-alignment and uniform response of all the continuous gas detectors used to generate a continuously combined/compensated concentration measurement signal. Gas analyzer systems must be optimized such that their overall response to rapid change in concentration is updated and recorded at an appropriate frequency to prevent loss of information. This test also verifies that the measurement system meets a minimum response time. For this procedure, ensure that all compensation algorithms and humidity corrections are turned on. You may use the results of this test to determine transformation time, t
(c) System requirements. Demonstrate that each continuously combined/compensated concentration measurement has adequate updating and recording frequencies and has a minimum rise time and a minimum fall time during a system response to a rapid change in multiple gas concentrations, including H
(1) The product of the mean rise time, t
(2) The frequency at which the system records an updated concentration must be at least 5 Hz. This criterion assumes that the frequency content of significant changes in emission concentrations during emission testing do not exceed 1 Hz. Also, the mean rise time must be at or below 10 seconds and the mean fall time must be at or below 10 seconds.
(3) You may use other criteria if we approve them in advance.
(4) You may meet the overall PEMS verification in § 1065.920 instead of the verification in this section for field testing with PEMS.
(d) Procedure. Use the following procedure to verify the response of each continuously compensated analyzer (verify the combined signal, not each individual continuously combined concentration signal):
(1) Instrument setup. Follow the analyzer manufacturer's start-up and operating instructions. Adjust the measurement system as needed to optimize performance. Run this verification with the analyzer operating in the same manner you will use for emission testing. If the analyzer shares its sampling system with other analyzers, and if gas flow to the other analyzers will affect the system response time, then start up and operate the other analyzers while running this verification test. You may run this verification test on multiple analyzers sharing the same sampling system at the same time. If you use any analog or real-time digital filters during emission testing, you must operate those filters in the same manner during this verification.
(2) Equipment setup. We recommend using minimal lengths of gas transfer lines between all connections and fast-acting three-way valves (2 inlets, 1 outlet) to control the flow of zero and blended span gases to the sample system's probe inlet or a tee near the outlet of the probe. If you inject the gas at a tee near the outlet of the probe, you may correct the transformation time, t
(3) Data collection. (i) Start the flow of zero gas.
(ii) Allow for stabilization, accounting for transport delays and the slowest analyzer's full response.
(iii) Start recording data. For this verification you must record data at a frequency greater than or equal to that of the updating-recording frequency used during emission testing. You may not use interpolation or filtering to alter the recorded values.
(iv) Switch the flow to allow the blended span gases to flow to the analyzer. If you intend to use the data from this test to determine t
(v) Allow for transport delays and the slowest analyzer's full response.
(vi) Switch the flow to allow zero gas to flow to the analyzer. If you intend to use the data from this test to determine t
(vii) Allow for transport delays and the slowest analyzer's full response.
(viii) Repeat the steps in paragraphs (d)(3)(iv) through (vii) of this section to record seven full cycles, ending with zero gas flowing to the analyzers.
(ix) Stop recording.
(e) Performance evaluations. (1) If you choose to demonstrate compliance with paragraph (c)(1) of this section, use the data from paragraph (d)(3) of this section to calculate the mean rise time, t
(2) If a measurement system fails the criterion in paragraph (e)(1) of this section, ensure that signals from the system are updated and recorded at a frequency of at least 5 Hz. In no case may the mean rise time or mean fall time be greater than 10 seconds.
(3) If a measurement system fails the criteria in paragraphs (e)(1) and (2) of this section, you may use the measurement system only if the deficiency does not adversely affect your ability to show compliance with the applicable standards in this chapter.
(f) Transformation time, t
(g) Optional procedure. Follow the optional procedures in § 1065.308(g), noting that you may use compensating gases mixed with ambient air for oxygen analyzers.
(h) Analyzers with H
(1) The analyzer is located downstream of a sample dryer.
(2) The maximum value for H
§ 1065.320 - Fuel-flow calibration.
(a) Calibrate fuel-flow meters upon initial installation. Follow the instrument manufacturer's instructions and use good engineering judgment to repeat the calibration.
(b) [Reserved]
(c) You may remove system components for off-site calibration. When installing a flow meter with an off-site calibration, we recommend that you consider the effects of the tubing configuration upstream and downstream of the flow meter. We recommend specifying calibration reference quantities that are NIST-traceable within ±0.5% uncertainty.
§ 1065.325 - Intake-flow calibration.
(a) Calibrate intake-air flow meters upon initial installation. Follow the instrument manufacturer's instructions and use good engineering judgment to repeat the calibration. We recommend using a calibration subsonic venturi, ultrasonic flow meter or laminar flow element. We recommend using calibration reference quantities that are NIST-traceable within ±0.5% uncertainty.
(b) You may remove system components for off-site calibration. When installing a flow meter with an off-site calibration, we recommend that you consider the effects of the tubing configuration upstream and downstream of the flow meter. We recommend specifying calibration reference quantities that are NIST-traceable within ±0.5% uncertainty.
(c) If you use a subsonic venturi or ultrasonic flow meter for intake flow measurement, we recommend that you calibrate it as described in § 1065.340.
§ 1065.330 - Exhaust-flow calibration.
(a) Calibrate exhaust-flow meters upon initial installation. Follow the instrument manufacturer's instructions and use good engineering judgment to repeat the calibration. We recommend that you use a calibration subsonic venturi or ultrasonic flow meter and simulate exhaust temperatures by incorporating a heat exchanger between the calibration meter and the exhaust-flow meter. If you can demonstrate that the flow meter to be calibrated is insensitive to exhaust temperatures, you may use other reference meters such as laminar flow elements, which are not commonly designed to withstand typical raw exhaust temperatures. We recommend using calibration reference quantities that are NIST-traceable within ±0.5% uncertainty.
(b) You may remove system components for off-site calibration. When installing a flow meter with an off-site calibration, we recommend that you consider the effects of the tubing configuration upstream and downstream of the flow meter. We recommend specifying calibration reference quantities that are NIST-traceable within ±0.5% uncertainty.
(c) If you use a subsonic venturi or ultrasonic flow meter for raw exhaust flow measurement, we recommend that you calibrate it as described in § 1065.340.
§ 1065.340 - Diluted exhaust flow (CVS) calibration.
(a) Overview. This section describes how to calibrate flow meters for diluted exhaust constant-volume sampling (CVS) systems.
(b) Scope and frequency. Perform this calibration while the flow meter is installed in its permanent position, except as allowed in paragraph (c) of this section. Perform this calibration after you change any part of the flow configuration upstream or downstream of the flow meter that may affect the flow-meter calibration. Perform this calibration upon initial CVS installation and whenever corrective action does not resolve a failure to meet the diluted exhaust flow verification (i.e., propane check) in § 1065.341.
(c) Ex-situ CFV and SSV calibration. You may remove a CFV or SSV from its permanent position for calibration as long as it meets the following requirements when installed in the CVS:
(1) Upon installation of the CFV or SSV into the CVS, use good engineering judgment to verify that you have not introduced any leaks between the CVS inlet and the venturi.
(2) After ex-situ venturi calibration, you must verify all venturi flow combinations for CFVs or at minimum of 10 flow points for an SSV using the propane check as described in § 1065.341. Your propane check result for each venturi flow point may not exceed the tolerance in § 1065.341(f)(5).
(3) To verify your ex-situ calibration for a CVS with more than a single CFV, perform the following check to verify that there are no flow meter entrance effects that can prevent you from passing this verification.
(i) Use a constant flow device like a CFO kit to deliver a constant flow of propane to the dilution tunnel.
(ii) Measure hydrocarbon concentrations at a minimum of 10 separate flow rates for an SSV flow meter, or at all possible flow combinations for a CFV flow meter, while keeping the flow of propane constant. We recommend selecting CVS flow rates in a random order.
(iii) Measure the concentration of hydrocarbon background in the dilution air at the beginning and end of this test. Subtract the average background concentration from each measurement at each flow point before performing the regression analysis in paragraph (c)(3)(iv) of this section.
(iv) Perform a power regression using all the paired values of flow rate and corrected concentration to obtain a relationship in the form of y = a · x
(d) Reference flow meter. Calibrate a CVS flow meter using a reference flow meter such as a subsonic venturi flow meter, a long-radius ASME/NIST flow nozzle, a smooth approach orifice, a laminar flow element, a set of critical flow venturis, or an ultrasonic flow meter. Use a reference flow meter that reports quantities that are NIST-traceable within ±1% uncertainty. Use this reference flow meter's response to flow as the reference value for CVS flow-meter calibration.
(e) Configuration. Calibrate the system with any upstream screens or other restrictions that will be used during testing and that could affect the flow ahead of the CVS flow meter, using good engineering judgment to minimize the effect on the flow distribution. You may not use any upstream screen or other restriction that could affect the flow ahead of the reference flow meter, unless the flow meter has been calibrated with such a restriction. In the case of a free standing SSV reference flow meter, you may not have any upstream screens.
(f) PDP calibration. Calibrate a positive-displacement pump (PDP) to determine a flow-versus-PDP speed equation that accounts for flow leakage across sealing surfaces in the PDP as a function of PDP inlet pressure. Determine unique equation coefficients for each speed at which you operate the PDP. Calibrate a PDP flow meter as follows:
(1) Connect the system as shown in Figure 1 of this section.
(2) Leaks between the calibration flow meter and the PDP must be less than 0.3% of the total flow at the lowest calibrated flow point; for example, at the highest restriction and lowest PDP-speed point.
(3) While the PDP operates, maintain a constant temperature at the PDP inlet within ±2% of the mean absolute inlet temperature, T
(4) Set the PDP speed to the first speed point at which you intend to calibrate.
(5) Set the variable restrictor to its wide-open position.
(6) Operate the PDP for at least 3 min to stabilize the system. Continue operating the PDP and record the mean values of at least 30 seconds of sampled data of each of the following quantities:
(i) The mean flow rate of the reference flow meter, n
(ii) The mean temperature at the PDP inlet, T
(iii) The mean static absolute pressure at the PDP inlet, p
(iv) The mean static absolute pressure at the PDP outlet, p
(v) The mean PDP speed, f
(7) Incrementally close the restrictor valve to decrease the absolute pressure at the inlet to the PDP, p
(8) Repeat the steps in paragraphs (e)(6) and (7) of this section to record data at a minimum of six restrictor positions ranging from the wide open restrictor position to the minimum expected pressure at the PDP inlet or the maximum expected differential (outlet minus inlet) pressure across the PDP during testing.
(9) Calibrate the PDP by using the collected data and the equations in § 1065.640.
(10) Repeat the steps in paragraphs (e)(6) through (9) of this section for each speed at which you operate the PDP.
(11) Use the equations in § 1065.642 to determine the PDP flow equation for emission testing.
(12) Verify the calibration by performing a CVS verification (i.e., propane check) as described in § 1065.341.
(13) During emission testing ensure that the PDP is not operated either below the lowest inlet pressure point or above the highest differential pressure point in the calibration data.
(g) SSV calibration. Calibrate a subsonic venturi (SSV) to determine its calibration coefficient, C
(1) Connect the system as shown in Figure 1 of this section.
(2) Verify that any leaks between the calibration flow meter and the SSV are less than 0.3% of the total flow at the highest restriction.
(3) Start the blower downstream of the SSV.
(4) While the SSV operates, maintain a constant temperature at the SSV inlet within ±2% of the mean absolute inlet temperature, T
(5) Set the variable restrictor or variable-speed blower to a flow rate greater than the greatest flow rate expected during testing. You may not extrapolate flow rates beyond calibrated values, so we recommend that you make sure the Reynolds number, Re
(6) Operate the SSV for at least 3 min to stabilize the system. Continue operating the SSV and record the mean of at least 30 seconds of sampled data of each of the following quantities:
(i) The mean flow rate of the reference flow meter n
(ii) Optionally, the mean dewpoint of the calibration air,T
(iii) The mean temperature at the venturi inlet,T
(iv) The mean static absolute pressure at the venturi inlet, P
(v) The mean static differential pressure between the static pressure at the venturi inlet and the static pressure at the venturi throat, ΔP
(7) Incrementally close the restrictor valve or decrease the blower speed to decrease the flow rate.
(8) Repeat the steps in paragraphs (g)(6) and (7) of this section to record data at a minimum of ten flow rates.
(9) Determine an equation to quantify C
(10) Verify the calibration by performing a CVS verification (i.e., propane check) as described in § 1065.341 using the new C
(11) Use the SSV only between the minimum and maximum calibrated Re
(12) Use the equations in § 1065.642 to determine SSV flow during a test.
(h) CFV calibration. Calibrate a critical-flow venturi (CFV) to verify its discharge coefficient, C
(1) Connect the system as shown in Figure 1 of this section.
(2) Verify that any leaks between the calibration flow meter and the CFV are less than 0.3% of the total flow at the highest restriction.
(3) Start the blower downstream of the CFV.
(4) While the CFV operates, maintain a constant temperature at the CFV inlet within ±2% of the mean absolute inlet temperature, T
(5) Set the variable restrictor to its wide-open position. Instead of a variable restrictor, you may alternately vary the pressure downstream of the CFV by varying blower speed or by introducing a controlled leak. Note that some blowers have limitations on nonloaded conditions.
(6) Operate the CFV for at least 3 min to stabilize the system. Continue operating the CFV and record the mean values of at least 30 seconds of sampled data of each of the following quantities:
(i) The mean flow rate of the reference flow meter, n
(ii) The mean dewpoint of the calibration air,T
(iii) The mean temperature at the venturi inlet,T
(iv) The mean static absolute pressure at the venturi inlet, P
(v) The mean static differential pressure between the CFV inlet and the CFV outlet, ΔP
(7) Incrementally close the restrictor valve or decrease the downstream pressure to decrease the differential pressure across the CFV, Δp
(8) Repeat the steps in paragraphs (f)(6) and (7) of this section to record mean data at a minimum of ten restrictor positions, such that you test the fullest practical range of ΔP
(9) Determine C
(10) Use C
(11) Verify the calibration by performing a CVS verification (i.e., propane check) as described in § 1065.341.
(12) If your CVS is configured to operate more than one CFV at a time in parallel, calibrate your CVS by one of the following:
(i) Calibrate every combination of CFVs according to this section and § 1065.640. Refer to § 1065.642 for instructions on calculating flow rates for this option.
(ii) Calibrate each CFV according to this section and § 1065.640. Refer to § 1065.642 for instructions on calculating flow rates for this option.
(i) Ultrasonic flow meter calibration. [Reserved]
§ 1065.341 - CVS and PFD flow verification (propane check).
This section describes two optional methods, using propane as a tracer gas, to verify CVS and PFD flow streams. You may use good engineering judgment and safe practices to use other tracer gases, such as CO
(a) A propane check uses either a reference mass or a reference flow rate of C
(b) Prepare for the propane check as follows:
(1) If you use a reference mass of C
(2) Select appropriate flow rates for the CVS and C
(3) Select a C
(4) Operate and stabilize the CVS.
(5) Preheat or pre-cool any heat exchangers in the sampling system.
(6) Allow heated and cooled components such as sample lines, filters, chillers, and pumps to stabilize at operating temperature.
(7) You may purge the HC sampling system during stabilization.
(8) If applicable, perform a vacuum side leak verification of the HC sampling system as described in § 1065.345.
(9) You may also conduct any other calibrations or verifications on equipment or analyzers.
(c) If you performed the vacuum-side leak verification of the HC sampling system as described in paragraph (b)(8) of this section, you may use the HC contamination procedure in § 1065.520(g) to verify HC contamination. Otherwise, zero, span, and verify contamination of the HC sampling system, as follows:
(1) Select the lowest HC analyzer range that can measure the C
(2) Zero the HC analyzer using zero air introduced at the analyzer port.
(3) Span the HC analyzer using C
(4) Overflow zero air at the HC probe inlet or into a tee near the outlet of the probe.
(5) Measure the stable HC concentration of the HC sampling system as overflow zero air flows. For batch HC measurement, fill the batch container (such as a bag) and measure the HC overflow concentration.
(6) If the overflow HC concentration exceeds 2 µmol/mol, do not proceed until contamination is eliminated. Determine the source of the contamination and take corrective action, such as cleaning the system or replacing contaminated portions.
(7) When the overflow HC concentration does not exceed 2 µmol/mol, record this value as x
(d) Perform the propane check as follows:
(1) For batch HC sampling, connect clean storage media, such as evacuated bags.
(2) Operate HC measurement instruments according to the instrument manufacturer's instructions.
(3) If you will correct for dilution air background concentrations of HC, measure and record background HC in the dilution air.
(4) Zero any integrating devices.
(5) Begin sampling, and start any flow integrators.
(6) Release the contents of the C
(7) Continue to release the cylinder's contents until at least enough C
(8) Shut off the C
(9) Stop sampling and stop any integrators.
(e) Perform post-test procedure as follows:
(1) If you used batch sampling, analyze batch samples as soon as practical.
(2) After analyzing HC, correct for contamination and background.
(3) Calculate total C
(4) If you use a reference mass, determine the cylinder's propane mass within ±0.5% and determine the C
(5) Subtract the reference C
(f) A failed propane check might indicate one or more problems requiring corrective action, as follows:
Table 1 of § 1065.341—Troubleshooting Guide for Propane Checks
Problem | Recommended corrective action |
---|---|
Incorrect analyzer calibration | Recalibrate, repair, or replace the FID analyzer. |
Leaks | Inspect CVS tunnel, connections, fasteners, and HC sampling system. Repair or replace components. |
Poor mixing | Perform the verification as described in this section while traversing a sampling probe across the tunnel's diameter, vertically and horizontally. If the analyzer response indicates any deviation exceeding ±2% of the mean measured concentration, consider operating the CVS at a higher flow rate or installing a mixing plate or orifice to improve mixing. |
Hydrocarbon contamination in the sample system | Perform the hydrocarbon-contamination verification as described in § 1065.520. |
Change in CVS calibration | Perform a calibration of the CVS flow meter as described in § 1065.340. |
Flow meter entrance effects | Inspect the CVS tunnel to determine whether the entrance effects from the piping configuration upstream of the flow meter adversely affect the flow measurement. |
Other problems with the CVS or sampling verification hardware or software | Inspect the CVS system and related verification hardware, and software for discrepancies. |
(g) You may verify flow measurements in a PFD (usually dilution air and diluted exhaust streams) for determining the dilution ratio in the PFD using the following method:
(1) Configure the HC sampling system to extract a sample from the PFD's diluted exhaust stream (such as near a PM filter). If the absolute pressure at this location is too low to extract an HC sample, you may sample HC from the PFD's pump exhaust. Use caution when sampling from pump exhaust because an otherwise acceptable pump leak downstream of a PFD diluted exhaust flow meter will cause a false failure of the propane check.
(2) Perform the propane check described in paragraphs (b), (c), and (d) of this section, but sample HC from the PFD's diluted exhaust stream. Inject the propane in the same exhaust stream that the PFD is sampling from (either CVS or raw exhaust stack).
(3) Calculate C
(4) Subtract the reference C
(h) Table 2 of § 1065.307 describes optional verification procedures you may perform instead of linearity verification for certain flow-measurement systems. Performing carbon balance error verification also replaces any required propane checks.
§ 1065.342 - Sample dryer verification.
(a) Scope and frequency. If you use a sample dryer as allowed in § 1065.145(e)(2) to remove water from the sample gas, verify the performance upon installation, after major maintenance, for thermal chiller. For osmotic membrane dryers, verify the performance upon installation, after major maintenance, and within 35 days of testing.
(b) Measurement principles. Water can inhibit an analyzer's ability to properly measure the exhaust component of interest and thus is sometimes removed before the sample gas reaches the analyzer. For example water can negatively interfere with a CLD's NO
(c) System requirements. The sample dryer must meet the specifications as determined in § 1065.145(e)(2) for dewpoint, T
(d) Sample dryer verification procedure. Use the following method to determine sample dryer performance. Run this verification with the dryer and associated sampling system operating in the same manner you will use for emission testing (including operation of sample pumps). You may run this verification test on multiple sample dryers sharing the same sampling system at the same time. You may run this verification on the sample dryer alone, but you must use the maximum gas flow rate expected during testing. You may use good engineering judgment to develop a different protocol.
(1) Use PTFE or stainless steel tubing to make necessary connections.
(2) Humidify room air, purified N
(3) Introduce the humidified gas upstream of the sample dryer. You may disconnect the transfer line from the probe and introduce the humidified gas at the inlet of the transfer line of the sample system used during testing. You may use the sample pumps in the sample system to draw gas through the vessel.
(4) Maintain the sample lines, fittings, and valves from the location where the humidified gas water content is measured to the inlet of the sampling system at a temperature at least 5 °C above the local humidified gas dewpoint. For dryers used in NO
(5) Measure the humidified gas dewpoint, T
(6) Measure the humidified gas dewpoint, T
(7) The sample dryer meets the verification if the dewpoint at the sample dryer pressure as measured in paragraph (d)(6) of this section is less than the dewpoint corresponding to the sample dryer specifications as determined in § 1065.145(e)(2) plus 2 °C or if the mole fraction of water as measured in (d)(6) is less than the corresponding sample dryer specifications plus 0.002 mol/mol.
(e) Alternate sample dryer verification procedure. The following method may be used in place of the sample dryer verification procedure in (d) of this section. If you use a humidity sensor for continuous monitoring of dewpoint at the sample dryer outlet you may skip the performance check in § 1065.342(d), but you must make sure that the dryer outlet humidity is at or below the minimum value used for quench, interference, and compensation checks.
§ 1065.345 - Vacuum-side leak verification.
(a) Scope and frequency. Verify that there are no significant vacuum-side leaks using one of the leak tests described in this section. For laboratory testing, perform the vacuum-side leak verification upon initial sampling system installation, within 8 hours before the start of the first test interval of each duty-cycle sequence, and after maintenance such as pre-filter changes. For field testing, perform the vacuum-side leak verification after each installation of the sampling system on the vehicle, prior to the start of the field test, and after maintenance such as pre-filter changes. This verification does not apply to any full-flow portion of a CVS dilution system.
(b) Measurement principles. A leak may be detected either by measuring a small amount of flow when there should be zero flow, or by detecting the dilution of a known concentration of span gas when it flows through the vacuum side of a sampling system.
(c) Low-flow leak test. Test a sampling system for low-flow leaks as follows:
(1) Seal the probe end of the system by taking one of the following steps:
(i) Cap or plug the end of the sample probe.
(ii) Disconnect the transfer line at the probe and cap or plug the transfer line.
(iii) Close a leak-tight valve located in the sample transfer line within 92 cm of the probe.
(2) Operate all vacuum pumps. After stabilizing, verify that the flow through the vacuum-side of the sampling system is less than 0.5% of the system's normal in-use flow rate. You may estimate typical analyzer and bypass flows as an approximation of the system's normal in-use flow rate.
(d) Dilution-of-span-gas leak test. You may use any gas analyzer for this test. If you use a FID for this test, correct for any HC contamination in the sampling system according to § 1065.660. If you use an O
(1) Prepare a gas analyzer as you would for emission testing.
(2) Supply reference gas to the analyzer span port and record the measured value.
(3) Route overflow reference gas to the inlet of the sample probe or at a tee fitting in the transfer line near the exit of the probe. You may use a valve upstream of the overflow fitting to prevent overflow of reference gas out of the inlet of the probe, but you must then provide an overflow vent in the overflow supply line.
(4) Verify that the measured overflow reference gas concentration is within ±0.5% of the concentration measured in paragraph (d)(2) of this section. A measured value lower than expected indicates a leak, but a value higher than expected may indicate a problem with the reference gas or the analyzer itself. A measured value higher than expected does not indicate a leak.
(e) Vacuum-decay leak test. To perform this test you must apply a vacuum to the vacuum-side volume of your sampling system and then observe the leak rate of your system as a decay in the applied vacuum. To perform this test you must know the vacuum-side volume of your sampling system to within ±10% of its true volume. For this test you must also use measurement instruments that meet the specifications of subpart C of this part and of this subpart D. Perform a vacuum-decay leak test as follows:
(1) Seal the probe end of the system as close to the probe opening as possible by taking one of the following steps:
(i) Cap or plug the end of the sample probe.
(ii) Disconnect the transfer line at the probe and cap or plug the transfer line.
(iii) Close a leak-tight valve located in the sample transfer line within 92 cm of the probe.
(2) Operate all vacuum pumps. Draw a vacuum that is representative of normal operating conditions. In the case of sample bags, we recommend that you repeat your normal sample bag pump-down procedure twice to minimize any trapped volumes.
(3) Turn off the sample pumps and seal the system. Measure and record the absolute pressure of the trapped gas and optionally the system absolute temperature. Wait long enough for any transients to settle and long enough for a leak at 0.5% to have caused a pressure change of at least 10 times the resolution of the pressure transducer, then again record the pressure and optionally temperature.
(4) Calculate the leak flow rate based on an assumed value of zero for pumped-down bag volumes and based on known values for the sample system volume, the initial and final pressures, optional temperatures, and elapsed time. Using the calculations specified in § 1065.644, verify that the vacuum-decay leak flow rate is less than 0.5% of the system's normal in-use flow rate.
§ 1065.350 - H2O interference verification for CO2 NDIR analyzers.
(a) Scope and frequency. If you measure CO
(b) Measurement principles. H
(c) System requirements. A CO
(d) Procedure. Perform the interference verification as follows:
(1) Start, operate, zero, and span the CO
(2) Create a humidified test gas by bubbling zero gas that meets the specifications in § 1065.750 through distilled H
(3) Introduce the humidified test gas into the sample system. You may introduce it downstream of any sample dryer, if one is used during testing.
(4) If the sample is not passed through a dryer during this verification test, measure the H
(5) If a sample dryer is not used in this verification test, use good engineering judgment to prevent condensation in the transfer lines, fittings, or valves from the point where x
(6) Allow time for the analyzer response to stabilize. Stabilization time may include time to purge the transfer line and to account for analyzer response.
(7) Operate the analyzer to get a reading for CO
(8) The analyzer meets the interference verification if the result of paragraph (d)(7) of this section meets the tolerance in paragraph (c) of this section.
(e) Exceptions. The following exceptions apply:
(1) You may omit this verification if you can show by engineering analysis that for your CO
(2) You may use a CO
§ 1065.355 - H2O and CO2 interference verification for CO NDIR analyzers.
(a) Scope and frequency. If you measure CO using an NDIR analyzer, verify the amount of H
(b) Measurement principles. H
(c) System requirements. A CO NDIR analyzer must have combined H
(d) Procedure. Perform the interference verification as follows:
(1) Start, operate, zero, and span the CO NDIR analyzer as you would before an emission test. If the sample is passed through a dryer during emission testing, you may run this verification test with the dryer if it meets the requirements of § 1065.342. Operate the dryer at the same conditions as you will for an emission test. You may also run this verification test without the sample dryer.
(2) Create a humidified CO
(3) Introduce the humidified CO
(4) If the sample is not passed through a dryer during this verification test, measure the H
(5) If a sample dryer is not used in this verification test, use good engineering judgment to prevent condensation in the transfer lines, fittings, or valves from the point where x
(6) Allow time for the analyzer response to stabilize. Stabilization time may include time to purge the transfer line and to account for analyzer response.
(7) Operate the analyzer to get a reading for CO concentration and record results for 30 seconds. Calculate the arithmetic mean of this data.
(8) The analyzer meets the interference verification if the result of paragraph (d)(7) of this section meets the tolerance in paragraph (c) of this section.
(9) You may also run interference procedures for CO
(e) Exceptions. The following exceptions apply:
(1) You may omit this verification if you can show by engineering analysis that for your CO sampling system and your emission-calculation procedures, the combined CO
(2) You may use a CO NDIR analyzer that you determine does not meet this verification, as long as you try to correct the problem and the measurement deficiency does not adversely affect your ability to show that engines comply with all applicable emission standards.
§ 1065.357 - CO2 interference verification for H2O FTIR analyzers.
(a) Scope and frequency. If you measure H
(b) Measurement principles. CO
(c) System requirements. An H
(d) Procedure. Perform the interference verification as follows:
(1) Start, operate, zero, and span the H
(2) Use a CO
(3) Introduce the CO
(4) Allow time for the analyzer response to stabilize. Stabilization time may include time to purge the transfer line and to account for analyzer response.
(5) Operate the analyzer to get a reading for H
(6) The analyzer meets the interference verification if the result of paragraph (d)(5) of this section meets the tolerance in paragraph (c) of this section.
(e) Exceptions. The following exceptions apply:
(1) You may omit this verification for CO
(2) You may omit this verification if you can show by engineering analysis that for your H
(3) You may use an H
§ 1065.360 - FID optimization and verification.
(a) Scope and frequency. For all FID analyzers, calibrate the FID upon initial installation. Repeat the calibration as needed using good engineering judgment. For a FID that measures THC, perform the following steps:
(1) Optimize the response to various hydrocarbons after initial analyzer installation and after major maintenance as described in paragraph (c) of this section.
(2) Determine the methane (CH
(3) If you determine NMNEHC by subtracting from measured THC, determine the ethane (C
(4) You may determine the methane (CH
(b) Calibration. Use good engineering judgment to develop a calibration procedure, such as one based on the FID-analyzer manufacturer's instructions and recommended frequency for calibrating the FID. Alternately, you may remove system components for off-site calibration. For a FID that measures THC, calibrate using C
(c) THC FID response optimization. This procedure is only for FID analyzers that measure THC. Use good engineering judgment for initial instrument start-up and basic operating adjustment using FID fuel and zero air. Heated FIDs must be within their required operating temperature ranges. Optimize FID response at the most common analyzer range expected during emission testing. Optimization involves adjusting flows and pressures of FID fuel, burner air, and sample to minimize response variations to various hydrocarbon species in the exhaust. Use good engineering judgment to trade off peak FID response to propane calibration gases to achieve minimal response variations to different hydrocarbon species. For an example of trading off response to propane for relative responses to other hydrocarbon species, see SAE 770141 (incorporated by reference, see § 1065.1010). Determine the optimum flow rates and/or pressures for FID fuel, burner air, and sample and record them for future reference.
(d) THC FID CH
(1) Select a C
(2) Select a CH
(3) Start and operate the FID analyzer according to the manufacturer's instructions.
(4) Confirm that the FID analyzer has been calibrated using C
(5) Zero the FID with a zero gas that you use for emission testing.
(6) Span the FID with the C
(7) Introduce the CH
(8) Allow time for the analyzer response to stabilize. Stabilization time may include time to purge the analyzer and to account for its response.
(9) While the analyzer measures the CH
(10) For analyzers with multiple ranges, you need to perform the procedure in this paragraph (d) only on a single range.
(11) Divide the mean measured concentration by the recorded span concentration of the CH
(12) You may determine the response factor as a function of molar water concentration using the following procedures and use this response factor to account for the CH
(i) Humidify the CH
(ii) Divide each mean measured CH
(iii) Use the CH
(iv) Use this functional relationship to determine the response factor during an emission test.
(e) THC FID CH
(1) Perform a CH
(2) If RF
(3) If RF
(4) Determine a new RF
(5) For analyzers with multiple ranges, you need to perform the procedure in this paragraph (e) only on a single range.
(f) THC FID C
§ 1065.362 - Non-stoichiometric raw exhaust FID O2 interference verification.
(a) Scope and frequency. If you use FID analyzers for raw exhaust measurements from engines that operate in a non-stoichiometric mode of combustion (e.g., compression-ignition, lean-burn), verify the amount of FID O
(b) Measurement principles. Changes in O
(c) System requirements. Any FID analyzer used during testing must meet the FID O
(d) Procedure. Determine FID O
(1) Select three span reference gases that contain a C
(2) Confirm that the FID analyzer meets all the specifications of § 1065.360.
(3) Start and operate the FID analyzer as you would before an emission test. Regardless of the FID burner's air source during testing, use zero air as the FID burner's air source for this verification.
(4) Zero the FID analyzer using the zero gas used during emission testing.
(5) Span the FID analyzer using a span gas that you use during emission testing.
(6) Check the zero response of the FID analyzer using the zero gas used during emission testing. If the mean zero response of 30 seconds of sampled data is within ±0.5% of the span reference value used in paragraph (d)(5) of this section, then proceed to the next step; otherwise restart the procedure at paragraph (d)(4) of this section.
(7) Check the analyzer response using the span gas that has the minimum concentration of O
(8) Check the zero response of the FID analyzer using the zero gas used during emission testing. If the mean zero response of 30 seconds of stabilized sample data is within ±0.5% of the span reference value used in paragraph (d)(5) of this section, then proceed to the next step; otherwise restart the procedure at paragraph (d)(4) of this section.
(9) Check the analyzer response using the span gas that has the average concentration of O
(10) Check the zero response of the FID analyzer using the zero gas used during emission testing. If the mean zero response of 30 seconds of stabilized sample data is within ±0.5% of the span reference value used in paragraph (d)(5) of this section, proceed to the next step; otherwise restart the procedure at paragraph (d)(4) of this section.
(11) Check the analyzer response using the span gas that has the maximum concentration of O
(12) Check the zero response of the FID analyzer using the zero gas used during emission testing. If the mean zero response of 30 seconds of stabilized sample data is within ±0.5% of the span reference value used in paragraph (d)(5) of this section, then proceed to the next step; otherwise restart the procedure at paragraph (d)(4) of this section.
(13) Calculate the percent difference between x
(14) If the O
(i) Repeat the verification to determine if a mistake was made during the procedure.
(ii) Select zero and span gases for emission testing that contain higher or lower O
(iii) Adjust FID burner air, fuel, and sample flow rates. Note that if you adjust these flow rates on a THC FID to meet the O
(iv) Repair or replace the FID and repeat the O
(v) Demonstrate that the deficiency does not adversely affect your ability to demonstrate compliance with the applicable emission standards.
(15) For analyzers with multiple ranges, you need to perform the procedure in this paragraph (d) only on a single range.
§ 1065.365 - Nonmethane cutter penetration fractions and NMC FID response factors.
(a) Scope and frequency. If you use a FID analyzer and an NMC to measure methane (CH
(b) Measurement principles. An NMC is a heated catalyst that removes nonmethane hydrocarbons from an exhaust sample stream before the FID analyzer measures the remaining hydrocarbon concentration. An ideal NMC would have a CH
(c) System requirements. We do not require that you limit NMC penetration fractions to a certain range. However, we recommend that you optimize an NMC by adjusting its temperature to achieve a PF
(d) Procedure for a FID calibrated with the NMC. The following procedure describes the recommended method for verifying NMC performance and the required method for any gaseous-fueled engine, including dual-fuel and flexible-fuel engines.
(1) Select CH
(2) Start, operate, and optimize the NMC according to the manufacturer's instructions, including any temperature optimization.
(3) Confirm that the FID analyzer meets all the specifications of § 1065.360.
(4) Start and operate the FID analyzer according to the manufacturer's instructions.
(5) Zero and span the FID with the NMC as you would during emission testing. Span the FID through the NMC by using CH
(6) Introduce the C
(7) Allow time for the analyzer response to stabilize. Stabilization time may include time to purge the NMC and to account for the analyzer's response.
(8) While the analyzer measures a stable concentration, record 30 seconds of sampled data. Calculate the arithmetic mean of the analytical gas mixture.
(9) Calculate a reference concentration of C
(10) For any gaseous-fueled engine, including dual-fuel and flexible-fuel engines, repeat the steps in paragraphs (d)(6) through (9) of this section, but with the CH
(11) Use RFPF
(e) Procedure for a FID calibrated with propane, bypassing the NMC. If you use a single FID for THC and CH
(1) Select CH
(2) Start and operate the NMC according to the manufacturer's instructions, including any temperature optimization.
(3) Confirm that the FID analyzer meets all the specifications of § 1065.360.
(4) Start and operate the FID analyzer according to the manufacturer's instructions.
(5) Zero and span the FID as you would during emission testing. Span the FID by bypassing the NMC and by using C
(6) Introduce the C
(7) Allow time for the analyzer response to stabilize. Stabilization time may include time to purge the NMC and to account for the analyzer's response.
(8) While the analyzer measures a stable concentration, record 30 seconds of sampled data. Calculate the arithmetic mean of the analytical gas mixture.
(9) Reroute the flow path to bypass the NMC, introduce the C
(10) Divide the mean C
(11) Repeat the steps in paragraphs (e)(6) through (10) of this section, but with the CH
(f) Procedure for a FID calibrated with CH
(1) Select CH
(2) Start and operate the NMC according to the manufacturer's instructions, including any temperature optimization.
(3) Confirm that the FID analyzer meets all the specifications of § 1065.360.
(4) Start and operate the FID analyzer according to the manufacturer's instructions.
(5) Zero and span the FID as you would during emission testing. Span the FID by bypassing the NMC and by using CH
(6) Introduce the C
(7) Allow time for the analyzer response to stabilize. Stabilization time may include time to purge the NMC and to account for the analyzer's response.
(8) While the analyzer measures a stable concentration, record 30 seconds of sampled data. Calculate the arithmetic mean of the analytical gas mixture.
(9) Divide the mean C
(10) Introduce the CH
(11) Allow time for the analyzer response to stabilize. Stabilization time may include time to purge the NMC and to account for the analyzer's response.
(12) While the analyzer measures a stable concentration, record 30 seconds of sampled data. Calculate the arithmetic mean of these data points.
(13) Reroute the flow path to bypass the NMC, introduce the CH
(14) Divide the mean CH
(g) Test gas humidification. If you are generating gas mixtures as a function of the molar water concentration in the raw or diluted exhaust according to paragraph (d) of this section, create a humidified test gas by bubbling the analytical gas mixture that meets the specifications in § 1065.750 through distilled H
(1) If the sample does not pass through a dryer during emission testing, generate at least five different H
(2) If the sample passes through a dryer during emission testing, humidify your test gas to an H
§ 1065.366 - Interference verification for FTIR analyzers.
(a) Scope and frequency. If you measure CH
(b) Measurement principles. Certain species can interfere with analyzers by causing a response similar to the target analyte. If the analyzer uses compensation algorithms that utilize measurements of other gases to meet this interference verification, a correct result depends on simultaneously conducting these other measurements to test the compensation algorithms during the analyzer interference verification.
(c) System requirements. An FTIR analyzer must have combined interference that is within ±2% of the flow-weighted mean concentration of CH
(d) Procedure. Perform the interference verification for an FTIR analyzer using the same procedure that applies for N
§ 1065.369 - H2O, CO, and CO2 interference verification for photoacoustic alcohol analyzers.
(a) Scope and frequency. If you measure ethanol or methanol using a photoacoustic analyzer, verify the amount of H
(b) Measurement principles. H
(c) System requirements. Photoacoustic analyzers must have combined interference that is within (0.0 ±0.5) µmol/mol. We strongly recommend a lower interference that is within (0.0 ±0.25) µmol/mol.
(d) Procedure. Perform the interference verification by following the procedure in § 1065.375(d), comparing the results to paragraph (c) of this section.
§ 1065.370 - CLD CO2 and H2O quench verification.
(a) Scope and frequency. If you use a CLD analyzer to measure NO
(b) Measurement principles. H
(c) System requirements. A CLD analyzer must have a combined H
(d) CO
(1) Use PTFE or stainless steel tubing to make necessary connections.
(2) Configure the gas divider such that nearly equal amounts of the span and diluent gases are blended with each other.
(3) If the CLD analyzer has an operating mode in which it detects NO-only, as opposed to total NO
(4) Use a CO
(5) Use an NO span gas that meets the specifications of § 1065.750 and a concentration that is approximately twice the maximum NO concentration expected during emission testing.
(6) Zero and span the CLD analyzer. Span the CLD analyzer with the NO span gas from paragraph (d)(5) of this section through the gas divider. Connect the NO span gas to the span port of the gas divider; connect a zero gas to the diluent port of the gas divider; use the same nominal blend ratio selected in paragraph (d)(2) of this section; and use the gas divider's output concentration of NO to span the CLD analyzer. Apply gas property corrections as necessary to ensure accurate gas division.
(7) Connect the CO
(8) Connect the NO span gas to the diluent port of the gas divider.
(9) While flowing NO and CO
(10) Measure the NO concentration downstream of the gas divider with the CLD analyzer. Allow time for the analyzer response to stabilize. Stabilization time may include time to purge the transfer line and to account for analyzer response. While the analyzer measures the sample's concentration, record the analyzer's output for 30 seconds. Calculate the arithmetic mean concentration from these data, x
(11) Calculate the actual NO concentration at the gas divider's outlet, x
(12) Use the values recorded according to this paragraph (d) and paragraph (e) of this section to calculate quench as described in § 1065.675.
(e) H
(1) Use PTFE or stainless steel tubing to make necessary connections.
(2) If the CLD analyzer has an operating mode in which it detects NO-only, as opposed to total NO
(3) Use an NO span gas that meets the specifications of § 1065.750 and a concentration that is near the maximum concentration expected during emission testing.
(4) Zero and span the CLD analyzer. Span the CLD analyzer with the NO span gas from paragraph (e)(3) of this section, record the span gas concentration as x
(5) Create a humidified NO span gas by bubbling a NO gas that meets the specifications in § 1065.750 through distilled H
(6) Introduce the humidified NO test gas into the sample system. You may introduce it upstream or downstream of any sample dryer that is used during emission testing. Note that the sample dryer must meet the sample dryer verification check in § 1065.342.
(7) Measure the mole fraction of H
(8) Use good engineering judgment to prevent condensation in the transfer lines, fittings, or valves from the point where x
(9) Measure the humidified NO span gas concentration with the CLD analyzer. Allow time for the analyzer response to stabilize. Stabilization time may include time to purge the transfer line and to account for analyzer response. While the analyzer measures the sample's concentration, record the analyzer's output for 30 seconds. Calculate the arithmetic mean of these data, x
(f) Corrective action. If the sum of the H
(g) Exceptions. The following exceptions apply:
(1) You may omit this verification if you can show by engineering analysis that for your NO
(2) You may use a NO
§ 1065.372 - NDUV analyzer HC and H2O interference verification.
(a) Scope and frequency. If you measure NO
(b) Measurement principles. Hydrocarbons and H
(c) System requirements. A NO
(d) Procedure. Perform the interference verification as follows:
(1) Start, operate, zero, and span the NO
(2) We recommend that you extract engine exhaust to perform this verification. Use a CLD that meets the specifications of subpart C of this part to quantify NO
(3) Upstream of any sample dryer, if one is used during testing, introduce the engine exhaust to the NDUV analyzer.
(4) Allow time for the analyzer response to stabilize. Stabilization time may include time to purge the transfer line and to account for analyzer response.
(5) While all analyzers measure the sample's concentration, record 30 seconds of sampled data, and calculate the arithmetic means for the three analyzers.
(6) Subtract the CLD mean from the NDUV mean.
(7) Multiply this difference by the ratio of the flow-weighted mean HC concentration expected at the standard to the HC concentration measured during the verification.
(8) The analyzer meets the interference verification of this section if the result of paragraph (d)(7) of this section meets the tolerance in paragraph (c) of this section.
(e) Exceptions. The following exceptions apply:
(1) You may omit this verification if you can show by engineering analysis that for your NO
(2) You may use a NO
§ 1065.375 - Interference verification for N2O analyzers.
(a) Scope and frequency. This section describes how to perform interference verification for certain analyzers as described in § 1065.275. Perform interference verification after initial analyzer installation and after major maintenance.
(b) Measurement principles. Certain species can positively interfere with analyzers by causing a response similar to N
(c) System requirements. Analyzers must have combined interference that is within (0.0 ±1.0) µmol/mol. We strongly recommend a lower interference that is within (0.0 ±0.5) µmol/mol.
(d) Procedure. Perform the interference verification as follows:
(1) Start, operate, zero, and span the N
(2) Create a humidified test gas by bubbling a multi component span gas that incorporates the target interference species and meets the specifications in § 1065.750 through distilled H
(3) Introduce the humidified interference test gas into the sample system upstream or downstream of any sample dryer, if one is used during testing.
(4) If the sample is not passed through a dryer during this verification test, measure the H
(5) If a sample dryer is not used in this verification test, use good engineering judgment to prevent condensation in the transfer lines, fittings, or valves from the point where x
(6) Allow time for the analyzer response to stabilize. Stabilization time may include time to purge the transfer line and to account for analyzer response.
(7) While the analyzer measures the sample's concentration, record its output for 30 seconds. Calculate the arithmetic mean of this data. When performed with all the gases simultaneously, this is the combined interference.
(8) The analyzer meets the interference verification if the result of paragraph (d)(7) of this section meets the tolerance in paragraph (c) of this section.
(9) You may also run interference procedures separately for individual interference species. If the concentrations of the interference species used are higher than the maximum levels expected during testing, you may scale down each observed interference value (the arithmetic mean of 30 second data described in paragraph (d)(7) of this section) by multiplying the observed interference by the ratio of the maximum expected concentration value to the actual value used during this procedure. You may run separate interference concentrations of H
§ 1065.376 - Chiller NO2 penetration.
(a) Scope and frequency. If you use a chiller to dry a sample upstream of a NO
(b) Measurement principles. A chiller removes H
(c) System requirements. A chiller must allow for measuring at least 95% of the total NO
(d) Procedure. Use the following procedure to verify chiller performance:
(1) Instrument setup. Follow the analyzer and chiller manufacturers' start-up and operating instructions. Adjust the analyzer and chiller as needed to optimize performance.
(2) Equipment setup and data collection. (i) Zero and span the total NO
(ii) Select an NO
(iii) Overflow this calibration gas at the gas sampling system's probe or overflow fitting. Allow for stabilization of the total NO
(iv) Calculate the mean of 30 seconds of recorded total NO
(v) Stop flowing the NO
(vi) Next saturate the sampling system by overflowing a dewpoint generator's output, set at a dewpoint of 50 °C, to the gas sampling system's probe or overflow fitting. Sample the dewpoint generator's output through the sampling system and chiller for at least 10 minutes until the chiller is expected to be removing a constant rate of H
(vii) Immediately switch back to overflowing the NO
(viii) Correct x
(3) Performance evaluation. If x
(e) Exceptions. The following exceptions apply:
(1) You may omit this verification if you can show by engineering analysis that for your NO
(2) You may use a chiller that you determine does not meet this verification, as long as you try to correct the problem and the measurement deficiency does not adversely affect your ability to show that engines comply with all applicable emission standards.
§ 1065.377 - Interference verification for NH3 analyzers.
(a) Scope and frequency. This section describes how to perform interference verification for certain analyzers as described in § 1065.277. Perform interference verification after initial analyzer installation and after major maintenance.
(b) Measurement principles. Certain compounds can positively interfere with analyzers by causing a response similar to NH
(c) System requirements. Analyzers must have combined interference that is within (0.0 ±2.0) µmol/mol.
(d) Procedure. Perform the interference verification as follows:
(1) Start, operate, zero, and span the NH
(2) Except as specified in paragraph (d)(9) of this section, select a multi-component span gas meeting the specification of § 1065.750 that incorporates the all the appropriate interference species. Use a humidity generator that meets the requirements in § 1065.750(a)(6) to humidify the span gas. If the sample does not pass through a dryer during emission testing, humidify your test gas to an H
(3) Introduce the humidified interference test gas into the sample system upstream or downstream of any sample dryer, if one is used during testing.
(4) If the sample does not pass through a dryer during this verification test, measure the H
(5) If the verification procedure does not include a sample dryer, use good engineering judgment to prevent condensation in the transfer lines, fittings, or valves between the point of
(6) Allow time for the analyzer response to stabilize. Stabilization time may include time to purge the transfer line and to account for analyzer response.
(7) Operate the analyzer to measures the sample's NH
(8) The analyzer meets the interference verification if the result of paragraph (d)(7) of this section meets the tolerance in paragraph (c) of this section.
(9) You may instead perform interference verification procedures separately for individual interference species. The interference verification specified in paragraph (c) of this section applies based on the sum of the interference values from separate interference species. If the concentration of any interference species used is higher than the maximum levels expected during testing, you may scale down each observed interference value by multiplying the observed interference value by the ratio of the maximum expected concentration value to the concentration in the span gas. You may run separate H
§ 1065.378 - NO2-to-NO converter conversion verification.
(a) Scope and frequency. If you use an analyzer that measures only NO to determine NO
(b) Measurement principles. An NO
(c) System requirements. An NO
(d) Procedure. Use the following procedure to verify the performance of a NO
(1) Instrument setup. Follow the analyzer and NO
(2) Equipment setup. Connect an ozonator's inlet to a zero-air or oxygen source and connect its outlet to one port of a three-way tee fitting. Connect an NO span gas to another port, and connect the NO
(3) Adjustments and data collection. Perform this check as follows:
(i) Set ozonator air off, turn ozonator power off, and set the analyzer to NO mode. Allow for stabilization, accounting only for transport delays and instrument response.
(ii) Use an NO concentration that is representative of the peak total NO
(iii) Turn on the ozonator O
(iv) Switch the ozonator on and adjust the ozone generation rate so the NO measured by the analyzer is 20 percent of x
(v) Switch the NO
(vi) Switch off the ozonator but maintain gas flow through the system. The NO
(vii) Turn off the ozonator O
(4) Performance evaluation. Calculate the efficiency of the NO
(5) If the result is less than 95%, repair or replace the NO
(e) Exceptions. The following exceptions apply:
(1) You may omit this verification if you can show by engineering analysis that for your NO
(2) You may use a converter that you determine does not meet this verification, as long as you try to correct the problem and the measurement deficiency does not adversely affect your ability to show that engines comply with all applicable emission standards.
(3) You may request to verify converter conversion efficiency using an NO
§ 1065.390 - PM balance verifications and weighing process verification.
(a) Scope and frequency. This section describes three verifications.
(1) Independent verification of PM balance performance within 370 days before weighing any filter.
(2) Zero and span the balance within 12 h before weighing any filter.
(3) Verify that the mass determination of reference filters before and after a filter weighing session are less than a specified tolerance.
(b) Independent verification. Have the balance manufacturer (or a representative approved by the balance manufacturer) verify the balance performance within 370 days of testing. Balances have internal weights that compensate for drift due to environmental changes. These internal weights must be verified as part of this independent verification with external, certified calibration weights that meet the specifications in § 1065.790.
(c) Zeroing and spanning. You must verify balance performance by zeroing and spanning it with at least one calibration weight. Also, any external weights you use must meet the specifications in § 1065.790. Any weights internal to the PM balance used for this verification must be verified as described in paragraph (b) of this section.
(1) Use a manual procedure in which you zero the balance and span the balance with at least one calibration weight. If you normally use mean values by repeating the weighing process to improve the accuracy and precision of PM measurements, use the same process to verify balance performance.
(2) You may use an automated procedure to verify balance performance. For example most balances have internal weights for automatically verifying balance performance.
(d) Reference sample weighing. Verify all mass readings during a weighing session by weighing reference PM sample media (e.g., filters) before and after a weighing session. A weighing session may be as short as desired, but no longer than 80 hours, and may include both pre-test and post-test mass readings. We recommend that weighing sessions be eight hours or less. Successive mass determinations of each reference PM sample media (e.g., filter) must return the same value within ±10 µg or ±10% of the net PM mass expected at the standard (if known), whichever is higher. If successive reference PM sample media (e.g., filter) weighing events fail this criterion, invalidate all individual test media (e.g., filter) mass readings occurring between the successive reference media (e.g., filter) mass determinations. You may reweigh these media (e.g., filter) in another weighing session. If you invalidate a pre-test media (e.g., filter) mass determination, that test interval is void. Perform this verification as follows:
(1) Keep at least two samples of unused PM sample media (e.g., filters) in the PM-stabilization environment. Use these as references. If you collect PM with filters, select unused filters of the same material and size for use as references. You may periodically replace references, using good engineering judgment.
(2) Stabilize references in the PM stabilization environment. Consider references stabilized if they have been in the PM-stabilization environment for a minimum of 30 min, and the PM-stabilization environment has been within the specifications of § 1065.190(d) for at least the preceding 60 min.
(3) Exercise the balance several times with a reference sample. We recommend weighing ten samples without recording the values.
(4) Zero and span the balance. Using good engineering judgment, place a test mass such as a calibration weight on the balance, then remove it. After spanning, confirm that the balance returns to a zero reading within the normal stabilization time.
(5) Weigh each of the reference media (e.g., filters) and record their masses. We recommend using substitution weighing as described in § 1065.590(j). If you normally use mean values by repeating the weighing process to improve the accuracy and precision of the reference media (e.g., filter) mass, you must use mean values of sample media (e.g., filter) masses.
(6) Record the balance environment dewpoint, ambient temperature, and atmospheric pressure.
(7) Use the recorded ambient conditions to correct results for buoyancy as described in § 1065.690. Record the buoyancy-corrected mass of each of the references.
(8) Subtract each reference media's (e.g., filter's) buoyancy-corrected reference mass from its previously measured and recorded buoyancy-corrected mass.
(9) If any of the reference filters' observed mass changes by more than that allowed under this paragraph, you must invalidate all PM mass determinations made since the last successful reference media (e.g. filter) mass validation. You may discard reference PM media (e.g. filters) if only one of the filter's mass changes by more than the allowable amount and you can positively identify a special cause for that filter's mass change that would not have affected other in-process filters. Thus, the validation can be considered a success. In this case, you do not have to include the contaminated reference media when determining compliance with paragraph (d)(10) of this section, but the affected reference filter must be immediately discarded and replaced prior to the next weighing session.
(10) If any of the reference masses change by more than that allowed under this paragraph (d), invalidate all PM results that were determined between the two times that the reference masses were determined. If you discarded reference PM sample media according to paragraph (d)(9) of this section, you must still have at least one reference mass difference that meets the criteria in this paragraph (d). Otherwise, you must invalidate all PM results that were determined between the two times that the reference media (e.g., filters) masses were determined.
§ 1065.395 - Inertial PM balance verifications.
This section describes how to verify the performance of an inertial PM balance.
(a) Independent verification. Have the balance manufacturer (or a representative approved by the balance manufacturer) verify the inertial balance performance within 370 days before testing.
(b) Other verifications. Perform other verifications using good engineering judgment and instrument manufacturer recommendations.